Slika je simbolična

Precision handheld instruments P700 series

Koda: 9852002*LLG
Proizvajalec: LLG
Precision handheld instruments P700 series
For temperature, moisture and airflow.
For highly accurate temperature measurements with USB PC online interface. (Optional: Windows SmartGraph software for graphical and tabular documentation, etc). Integral probe calibration for sensor tolerance compensation. Min.- Max.- Hold- and AVG- functions. The large LCD shows two measurements synchronously, even temperature differences can be displayed (P705, P755, P755-Log, P770).

Specifications
General (all models):
USB interface, Max-Min memory, Hold function, true average calculation over a freely selectable period, Auto-off function, measured quantities can be freely assigned to instrument channels, comprehensive calibration options (selectable 1-point, 2-point or 3-point calibration), large display with bar graph for simultaneous display of 2 measured values
P700/P705:
1/2 measurement channels (Pt100 4-wire probe, thermocouples type J, K, L, N, T, R, S), measurement ranges and accuracies: Pt100:-200 to +850°C ( ±0.1 °C -100 to +200°C, otherwise ±0.1% of measurement value)
thermocouples, DIN types (-200 to +1760°C)
P710/P715:
1/2 measurement channels (Thermocouples type K, J, L, N, T), measurement ranges and accuracies: -200 to +1370°C (±0.2 °C from 0 to +200 °C, ±0.5 °C to 1000 °C, otherwise ± 1.0 °C of measurement value)
Thermocouples acc. to EN 60584
P750/P755/P755-Log:
1/2 measurement channels (Pt100 4-wire probe, thermocouples type J, K, L, N, T, R, S, moisture, flow), measurement ranges and accuracies: Pt100: -200 to +850°C (±0.03°C -50 to +200°C, ±0.05°C -200 to +200°C, otherwise ±0.05% of measurement value), thermocouples, DIN type (types J, K, L, N, T 0 to 200 °C, ±0.3°C, +200 to +1000°C ±1.0 °C, otherwise ±1.5°C;
type R, S ±1.0 °C +0.1% of measurement value) humidity 0 to 100% R.H. ( ±1.5%R.H.), flow 0 to 40 m/s (±0.5% of measurement value +1.0% of final value)
P770:
2 measurement channels (Pt100, thermocouples type J, K, L, N, T, R, S, moisture, flow), measurement ranges and accuracies: Pt100: -200 to +850°C ( ±0.1 °C -100 to +200°C, otherwise ±0.1% of measurement value)
thermocouples, DIN types, humidity 0 to 100%R.H. ( ±1.5%R.H.), flow 0 to 40 m/s (±0.5% of measurement value)
Dimensions (L x W x D)/ weight: 200 x 85 x 40mm / 300g

P700-P755 for Ex-use Zone 1 and Zone 2 according to ATEX approval II 2 G Ex ib IIB T4 Gb available on request.
Manufacturer:DOSTMANN electronic GmbH
Article description 1:Precision hand-held meter P710
Article description 2:1-channel, w/o probe